HOME
COMPANY
PRODUCT
APPLICATION
NEWS
SERVICE
CONTACT
Dissolved Oxygen Analyzer

2021-01-2054

BACK TO LIST

Dissolved Oxygen Analyzer


Measurement principle

Dissolved oxygen is the content of gaseous molecular oxygen dissolved in water. The content of dissolved oxygen in water is closely related to the partial pressure of oxygen in the air and the temperature of water.

Fluorescence quenching method: the blue pulsating light beam emitted by the internal optical system of the sensor hits the fluorescent layer, and the marker "responds" (produces fluorescence) with pulsating red light. The duration and intensity of the excited response signal are directly related to oxygen The composition is related to the partial pressure.  (*Amperometric method: Oxygen molecules penetrate the gas permeable membrane at the front of the sensor and are reduced by the working electrode to generate a diffusion current proportional to the oxygen concentration.)

Application

Sewage treatment plant

▲Drinking water plant

Rivers and lakes

Fishing

Circulating water of power plant boiler

Fermentation tank product quality monitoring in the food and pharmaceutical industry

Feature

Smart digital MEMS sensor

Sensor self-diagnosis, proactively reminding maintenance and management

The sensor automatically completes all compensation and measurement algorithms

Real-time temperature, salinity and pressure compensation

The sensor data communication bus cable can reach 1000 meters

Quick response sensor

Sensor IP68 protection level

Fluorescence measurement sensor

Good robustness

No calibration, no film, no fluoride, no drift

Unlimited PH value, CO2, H2S, SO2 influence

High-precision measurement under low dissolved oxygen concentration

No oxygen consumption

Comply with ASTM standard D888-05

Installation

No need to wait for polarization, measure immediately!

No need to replace the membrane and fluorocarbon, saving cost!

No need to disassemble, convenient!

No regeneration, it is impossible to predict the decay from sewage on the surface of the thermometer

H2S or ammonia in the liquid does not affect the measurement

No need to replace the film and carbon dioxide

No need to continuously perform on-site calibration

Clean the sensor surface: every month

Sensor calibration: every quarter

Replace the fluorescent cap: about every 2 years


Application Case

Sewage aeration tank: Nitrogen and phosphorus removal, COD degradation, and dissolved oxygen as an important parameter for controlling aeration.

Ion: Dissolved oxygen detection during iron and manganese ion removal process

Power plant: boiler water supply


Technical Parameters

Measurement range

0~20mg/L,0~200ppm,0~200%SAT

Accuracy

0.1%/0.1°C

Sensitivity

0.01mg/L/ 1%SAT

Response time(T90)

< 1 s

Automation Compersation

Atmospheric pressure,Salinity,Temperature

Liquid temperature

0~50℃

Pressure

Max. 5Bar

Ambient temperature

-10~50°C

Humility

0~90%

Sensor Size

Φ25mm*150mm

Material

SS316L/Titanium(sea water)


Controller features

• Quick and convenient

    The navigation menu contains 6 languages, which can be operated easily.

    Uses M12 connector to connect with sensor, Plug-and-play.

• Process safety

    4.3" or 7" large size color LCD touch screen, convenient and safe touch operation and debugging

    Large size screen with red flashing alarm, clearly visible from long distances and in dark areas

    Alarm immediately, safe the process

    Using M12 connector to connect with sensor, no risk of wrong connection

• Data alarm record

    Real-time data curve display

    Record function for up to 6000 alarms

• Expert calibration function

   Multi-point calibration function up to 9 point

• Powerful self-diagnosis function

    Built-in heartbeat monitoring function and watchdog

    Monitor the status of controllers and sensors, and promptly remind customers to take necessary maintenance

    High-standard hardware and software security and password protection

• Powerful control function

    High(low) limit control function

    Optional: Timer control(automatic cleaning) function

    Optional: analog PID controller

    Optional: PWM controller

• Flexible fieldbus communication functions for IOT4.0

    Optional fieldbus MODBUS, HART, Foundation Fieldbus FF, PROFIBUS PA, PROFIBUS DP, etc.

现场


Technical parameters (Controller)

Display

4.3" or 7" industrial color touch screen  

Language

6 languages

Diagnosis function

Sensor self-diagnosis

Heartbeat monitoring

6000 alarm record

Analog output

4~20mA,maximum load 500Ω

Alarm

Relay(2A), freely set alarm

Control function

Optional Timer controller

Optional PID analog controller

Optional PWM controller

Calibration

Expert calibration function

Multi-point calibration function up to 9 point

Communication

RS485 MODBUS RTU, HART,

Foundation Fieldbus FF,

PROFIBUS PA, PROFIBUS DP,

MODBUS TCP/IP, etc

Power

80~264V AC,1A

or 19~28V DC,3A

Ambient temperature

-10~55℃

Wall-mounted

(1~2Channels)

4.3" color touchscreen

213*185*84mm

7" color touchscreen

323x237x172mm

IP65,option Ex d IICT6

Portable

(1~2Channels)

(Built-in lithium battery)

7" color touchscreen

IP40 250x144x184mm

IP40 357x144x184mm

IP67 420x325x180mm

19" Rack

(1~6Channels)

7" color touchscreen

IP40 483x133x238mm




This website uses cookies. Further information can be found in our Privacy policy.